- Dimensions
- 330 × 126 × 110 mm
- Syringe size
- 5, 10, 20, 30, 50 or 60 ml
- Infusion mode
- Rate, Time and Body weight modes
- Flow rate
- 5 ml syringe: (0.1-150) ml/h; 10 ml syringe: (0.1-300) ml/h; 20 ml syringe: (0.1-600) ml/h; 30 ml syringe: (0.1-900) ml/h; 50 or 60 ml syringe: (0.1-1500) ml/h
- Total volume
- 0.1-9999 ml in 0.1 ml increments
- Volume to be infused (VTBI)
- 0.1-9999 ml in 0.1 ml increments
- Increment
- 0.1-999.9 ml/h: 0.1 ml/h increments; 1000-1500 ml/h: 1 ml/h increments
- Accuracy
- flow rate ≥1 ml/h: accuracy ≤± 2 %; flow rate <1 ml/h: accuracy ≤± 3 %
- Keep vein open (KVO) rate
- 0.1-5.0 ml/h, adjustable
- Bolus and purge rate
- 0.10-1500 ml/h (depending on syringe size)
- Occlusion
- low level: 300 ± 75 mmHg (40 kPa); medium level: 525 ± 75 mmHg (70 kPa); high level: 900 ± 135 mmHg (120 kPa)
- Display
- 3.0-inch monochrome LCD, 240×128 pixels
- Titration
- yes
- Keypad lock
- yes
- Pause
- yes
- Automatic syringe size detection
- yes
- Self-diagnostic system
- yes
- Last-settings configuration
- yes
- Drug library
- 200 drugs
- Event log
- 1500 events
- Weight
- <1.8 kg (including clamp)
- Docking station
- supported
- Handle
- included
- Mounting
- standard on a vertical pole, optional on a horizontal rail
- Alarm volume
- 1-8 selectable levels
- Preset syringe types
- 8 types, including B-D, B.Braun, Terumo and 5 Chinese brands; local configurations can be used
- Language
- Russian, English, Bulgarian, Chinese, Polish, Portuguese, Romanian, Turkish, French and Czech
- Alarms
- VTBI completed, VTBI near completion, KVO finished, syringe empty, syringe near empty, syringe disengaged, occlusion, system error, system malfunction, low battery, battery depleted, reminder, AC power disconnected, standby