- Flow rate
- 0.1-1500 ml/h (step 0.1 ml/h; above 100 ml/h, step 1 ml/h)
- Bolus rate
- 600-1000 ml/h (adjustable)
- Keep vein open (KVO) flow
- 0.1-5 ml/h (adjustable)
- Occlusion pressure levels
- H: 900 ± 200 mmHg (120.7 ± 26.7 kPa); C: 500 ± 100 mmHg (66.7 ± 13.3 kPa); L: 100 ± 50 mmHg (13.3 ± 6.6 kPa)
- Infusion accuracy
- no more than ±5% (no more than ±3% with calibrated infusion sets)
- Air bubble detector
- ultrasonic detection method, sensitivity 25 µl
- Number of infusion records
- 1500
- Preset volume
- 1-9999 ml
- Delivered volume display range
- 0.1-9999 ml (step 0.1 ml; above 100 ml, step 1 ml)
- Alarms
- occlusion; air bubble; end of infusion; door open; low battery; battery depleted; power failure; not calibrated; system error; standby; KVO mode completed; drop sensor failure
- Power supply
- AC 110-230 V, 50-60 Hz
- Internal battery
- DC 12 V, 2300 mAh; on a full charge, more than 6 hours of operation at 25 ml/h (including low battery time); higher capacity batteries can be used
- External power supply
- DC 12 V
- Operating conditions
- ambient temperature 5 to 40 °C, relative humidity 20% to 90%, ambient pressure 86-106 kPa
- Storage conditions
- ambient temperature -20 to +55 °C, relative humidity no more than 95%
- Electric shock protection
- class 1, internal power supply; type CF; continuous operation mode
- Water resistance
- IPX4
- Infusion modes
- flow mode; drip mode; weight mode and time mode; two groups of infusion programs can be set in flow and drip modes
- Infusion rate
- 3 modes: drops/min; ml/h; time/volume
- Weight
- 2.5 kg (with mounting base)
- Dimensions
- 308 mm (width) × 140 mm (height) × 144 mm (length)
- Infusion sets
- stores the settings of sets from 9 brands; infusion accuracy reaches ±3% after calibration
- Possible applications
- in addition to the pressure control performed by the pump, the flow rate and volume of fluid (drugs, nutrition) delivered to the patient can be controlled
- Contraindication
- not intended for blood transfusion