Strep A rapid test
Cassette rapid test for qualitative detection of group A streptococcal antigens in throat swabs; reading time 5 minutes, 20 tests per kit.
Price on request
Add this device to your quote list to request its price.ТОО "FSA KZ (ФСА КЗ)", Almaty, Kazakhstan, BIN 210940036794
- Brand
- STA
- Category
- Rapid tests
Description
Intended use
The Strep A rapid test (swab) is a rapid visual immunoassay for the qualitative presumptive detection of group A streptococcal antigens in human throat swab specimens. The Strep A rapid test is intended to aid in the diagnosis of infection caused by group A streptococcus.
Materials
- Individually packed test devices
- Reagent 1
- Reagent 2
- Positive control
- Negative control
- Sterile swabs
- Disposable pipettes
- Extraction tubes
- Workstation
- Package insert
Product features
STA: Strep A swab (cassette)
One-step in vitro diagnostic test based on immunochromatography
Specimen: swab
Reading time: 5 minutes
Storage: 2-30 °C
Kit pack: 20 pcs
Procedure
Bring tests, specimens, reagents and/or controls to room temperature (15-30 °C) before use.
1. Prepare the swab specimens:
- Place a clean extraction tube in the designated position of the workstation. Add 4 drops of reagent 1 to the extraction tube, then add 4 drops of reagent 2. Mix the solution by gently swirling the extraction tube.
- Immediately immerse the swab in the extraction tube. Rotate the swab in a circular motion against the wall of the extraction tube so that the liquid is expressed from the swab and can be reabsorbed.
- Let stand for 1-2 minutes at room temperature, then press the swab firmly against the tube to express as much liquid as possible from the swab. Discard the swab following the guidelines for handling infectious agents.
2. Remove the test from the sealed pouch and place it on a clean, level surface. Label the device with patient or control identification. For best results, the assay should be performed within one hour.
3. Using a disposable pipette, add 3 drops (approximately 120 µl) of the extracted solution from the extraction tube to the specimen well of the test device.
Avoid trapping air bubbles in the specimen well (S) and do not add solution to the result window.
As the test begins to work, color will migrate across the membrane.
4. Wait for the colored lines to appear. The result should be read at 5 minutes. Do not interpret the result after 10 minutes.
Similar devices
Whole category