- Analytical system
- fully automated random-access biochemistry system with STAT sample loading
- Analytical principle
- spectrophotometry and potentiometry
- Assay types
- end point, kinetic, fixed time and ISE (indirect measurement)
- Analytical methods
- colorimetry, turbidimetry, latex agglutination, homogeneous enzyme immunoassay, indirect ISE
- Simultaneous analyte processing
- 60 photometric tests + 3 ISE, 120 on-board preprogrammed tests
- Throughput
- 400 photometric tests per hour; maximum 800 with ISE
- Ion-selective ISE module throughput
- 200 samples per hour. Maximum tests per hour: 600 (ISE only)
- Sample types
- serum, plasma, urine and others
- Sample transport
- racks holding 10 samples each (barcodes on primary tubes and racks); capacity of 80 samples; continuous loading
- Sample tubes
- primary and secondary tubes; diameter 11.5-16 mm; height 55-102 mm, built-in micro sample cups.
- STAT samples
- 22 STAT sample positions on a refrigerated STAT rotor. Easy interruption between tests. Automatic rerun capability.
- Sample volume
- 1-25 µl in 0.1 µl steps (1-25 µl for reruns)
- Reagent capacity
- 76 positions (R1+R2, detergent position). Holds 15, 30 and 60 ml bottles.
- Reagent storage
- refrigerated (4-12 °C)
- Reagent volume
- R1: 10-250 µl; R2: 10-250 µl (in 1 µl steps)
- Total reaction volume
- 90-350 µl
- Reaction cuvette
- quartz cuvettes with unlimited service life.
- Reaction time
- up to 8 minutes 37.5 seconds.
- Reaction incubation
- 37 °C, dry bath.
- Photometric system
- direct measurement through the reaction cuvettes (optical density 0-3.0), monochromatic and bichromatic measurements possible.
- Wavelength
- 13 wavelengths from 340 to 800 nm
- Calibration
- automatic calibration, on-board refrigeration of calibrators; master calibration specified by two-dimensional barcode.
- Quality control
- automatic quality control, refrigerated quality control positions.
- Rerun
- 1) automatic and manual rerun; 2) automatic rerun with unchanged, reduced or increased sample volume; 3) reflex testing.
- Sample predilution
- fixed dilution ratios of 3, 5, 10, 15, 20, 25, 50, 76 and 100.
- Test requests
- online checking of individual test and profile orders using a mouse, function keys or touch screen.
- Sample integrity
- lipemia, hemolysis and icterus analysis; clot detection and protection against pipettor failures.
- Network settings
- RS232C. Full unidirectional and bidirectional communication possible.
- Software
- Windows 7®
- Data storage
- 100 000 samples.
- Reaction data
- 200 000 tests.
- Dimensions (W x H x D), analyzer
- 1 450 x 1 205 x 760
- Power supply
- 100-240 V; 60 Hz, <3.5 kVA