Urine analysis reagent strips (10 parameters)
In vitro diagnostic reagent strips for qualitative and semi-quantitative urinalysis; 100 strips per bottle, visual or instrument reading.
Price on request
Add this device to your quote list to request its price.ТОО "FSA KZ (ФСА КЗ)", Almaty, Kazakhstan, BIN 210940036794
- Category
- Rapid tests
Description
100 strips/bottle
Intended use
Urine analysis reagent strips are in vitro diagnostic reagents intended for qualitative and semi-quantitative urinalysis. The strips test for leukocytes, nitrite, urobilinogen, protein, pH, blood, specific gravity, ascorbic acid, ketone, bilirubin, glucose and microalbumin in urine. The specific test parameters of the product you are using are indicated on the outside of the box and on the bottle label.
Please read these instructions carefully before using the urine analysis reagent strips.
The results on the strips can be read visually or with an instrument.
Specifications
100 strips/bottle
Specimen collection and preparation
Use only a clean, dry container to collect urine; the test should be performed within 2 hours. All procedures should be carried out in a sanitary environment.
Caution
Water cannot be used as a negative quality control fluid. Urine preservatives cannot prevent the degradation of ketones, bilirubin and urobilinogen. If a urine specimen stands for a long time, the glucose, pH, nitrite and blood results may be affected by bacterial growth.
Test procedure
1. Remove one strip from the bottle and close the cap immediately.
2. Dip the reagent area of the strip into the urine specimen and remove it quickly.
3. Wipe excess urine off against the edge of the specimen container.
4. Read the results carefully within 60 seconds in good light, holding the test area next to the corresponding color chart on the bottle label. Color changes that appear only along the edges of the test pads or after more than 2 minutes have no diagnostic value. The leukocyte test pad can be read within 120 seconds.
For instrument reading, carefully follow the directions given in the operating manual of the instrument.